[SENSe[1]]|SENSe2:MRATe <character_data>
This command sets the measurement speed on the selected channel.
When a channel is set to FAST, the following couplings occur:
[SENSe[1]]|SENSe2:AVERage:STATe
[SENSe[1]]|SENSe2:CORRection:DCYCle:STATe
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe
CALCulate[1|2|3|4]:GAIN:STATe
CALCulate[1|2|3|4]:RELative:STATe
CALCulate1|3:MATH:EXPRession
CALCulate2|4:MATH:EXPRession
1. This change only occurs on the channel specified in the SENSe:MRATe
command. When the specified channel is changed from FAST to NORMal or
DOUBle, the settings that were in place when FAST was entered are restored.
2. This change occurs when either channel is set to FAST. When both channels are
changed from FAST to NORMal or DOUBle, the settings that were in place when
FAST was entered are restored.
3. Applicable to the E4417A dual channel power meter only.
Note
FAST mode cannot be used over RS232 or RS422 serial interfaces.
Syntax
SENS
SENS2
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
[SENSe[1]]|SENSe2:MRATe <character_data>
Command
:
MRAT
Space
1
SENSe Subsystem
Status
1
OFF
1
OFF
1
OFF
2
OFF
2
OFF
"(SENSe1)"
3
"(SENSe2)"
numeric_value
?