SERVice:BIST:FPATH[1]|2:STATe <boolean>
SERVice:BIST:FPATH[1]|2:STATe <boolean>
This command enables/disables fast path accuracy test mode for the
specified channel. It applies to E9320 Series power sensors only.
Note
Additional hardware is needed to run this command. For further
information refer to the power meter Service Guide.
After using this command, a system preset is recommended.
If the command is set to:
• ON, fast path accuracy test mode is enabled for the specified channel.
• OFF, fast path accuracy test mode is disabled for the specified channel.
Syntax
SERV
Example
SERV:BIST:FPATH:STATe OFF
Reset Condition
On reset, test mode is disabled.
Error Messages
• If no power sensor is connected, error -241 "Hardware missing" occurs.
• If a non E9320 Series power sensor is connected, error -241
"Hardware missing" occurs.
• If FPATH2 is sent to a single channel meter, error -113
"Undefined header" occurs.