INITiate[1]|2[:IMMediate]

This command sets the power meter in the wait for trigger state. When a
trigger is received, the measurement is taken and the result placed in the
power meter memory. If TRIGger:SOURce is set to IMMediate the
measurement begins as soon as INITiate:IMMediate is executed.
Use FETCh? to transfer a measurement from memory to the output buffer.
Refer to "FETCh[1]|2|3|4 Queries", on page 2-23 for further details.
Note
This command performs the same function as
INITiate:[IMMediate]:SEQuence[1]|2.
Syntax
Example
INIT2:IMM
Error Messages
If the power meter is not in the idle state or INITiate:CONTinuous is
ON, error -213, "INIT ignored" occurs.
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
INIT
1
:IMM
2
TRIGger Subsystem
INITiate[1]|2[:IMMediate]
This command places channel B in
the wait for trigger state.
13-7