SYSTem:COMMunicate:SERial:TRANsmit:PACE XON|NONE
This command enables (XON) and disables (NONE) an Xon/Xoff software
handshake for the transmitter. When enabled, if an Xoff control character
is detected by the receiver, the transmitter will not transmit further
characters until an Xon control character is detected by the receiver.
Syntax
SYST
Example
SYST:COMM:SER:TRAN:PACE XON
Reset Condition
On reset, pacing is unaffected.
Query
SYSTem:COMMunicate:SERial:TRANsmit:PACE?
The query returns XON if the handshake is enabled and NONE if the
handshake is disabled.
Query Example
SYST:COMM:SER:TRAN:PACE?
Agilent E4416A/E4417A Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SYSTem:COMMunicate:SERial:TRANsmit:PACE XON|NONE
:COMM
:SER
:TRAN
SYSTem Subsystem
Space
:PACE
XON
NONE
?
This command enables the
transmitter Xon/Xoff software
handshake.
This command queries whether
the transmitter Xon/Xoff software
handshake is enabled (XON) or
disabled (NONE).