SERVice Subsystem
SERVice:SENSor[1]|2:CALFactor <cal_factor_data>
Error Messages
• If no power sensor is connected, error -241 "Hardware missing" occurs.
• If a non E-Series sensor or N8480 Series sensor with Option CFT is
connected, error -241 "Hardware missing" occurs.
• If an E9320 Series sensor is connected and
SERVice:SENSor[1]|2:CORRections:STATe is set to ON, error -221,
"Settings conflict" occurs.
• If INIT:CONT is not set to OFF, error -221, "Settings conflict" occurs.