Memory:table:move <Character_Data>,<Character_Data> - Agilent Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

MEMory Subsystem
MEMory:TABLe:MOVE <character_data>,<character_data>
MEMory:TABLe:MOVE <character_data>,<character_data>
This command is used to rename a sensor calibration table (8480 Series
power sensors and N8480 Series sensors with Option CFT only) or a
frequency dependent offset table.
Syntax
:TABL
MEM
Parameters
Item
character_data
1st parameter)
character_data(
2nd parameter)
Example
MEM:TABL:MOVE "tab1","tab1a"
Error Messages
If either table name is invalid, error -224, "Illegal parameter value"
occurs.
If the first parameter does not match an existing table name, error
-256, "File name not found" occurs.
If the second parameter matches an existing table name or save/recall
register, error -257, "File name error" occurs.
7-30
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Space
character_data
:MOVE
Description/Default
Contains the existing table name.
Details the new table name. A
maximum of 12 characters can be
used.
EPM-P Series Power Meters Programming Guide
,
character_data
Range of Values
existing table
name
A to Z (uppercase)
a to z (lowercase)
0 - 9
_ (underscore)
This command renames a table
named tab1 to tab1a.