Calibration[1]|2:Rcalibration <Boolean> - Agilent Technologies EPM-P Series Programming Manual

Power meters
Hide thumbs Also See for EPM-P Series:
Table of Contents

Advertisement

CALibration Subsystem
CALibration[1]|2:RCALibration <boolean>
CALibration[1]|2:RCALibration <boolean>
This command enables and disables the zero/cal lockout facility. With the
lockout facility enabled the power meter is prevented from making
measurements until the sensor connected has been zeroed and calibrated.
Syntax
CAL
Example
CAL1:RCAL 1
Reset Condition
On reset, the state of the zero/cal lockout is unaffected.
Query
CALibration[1]|2:RCALibration?
The query enters a 1 or 0 into the output buffer indicating whether
zero/cal lockout is enabled or disabled.
Query Example
CAL1:RCAL?
4-10
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
1
:RCAL
2
1 is returned if zero/cal lockout is enabled.
0 is returned if zero/cal lockout is disabled.
EPM-P Series Power Meters Programming Guide
Space
0|OFF
1|ON
?
This command enables the zero/cal
lockout facility on channel A.
This command queries whether or
not the zero/cal lockout facility is
enabled for channel A.