Note
For most applications, automatic filter length selection
([SENSe[1]]|SENSe2:AVERage:COUNt:AUTO ON) is the best mode
of operation. However, manual filter length selection
([SENSe[1]]|SENSe2:AVERage:COUNt <numeric_value>) is
useful in applications requiring either high resolution or fast settling
times.
Syntax
SENS
1
SENS2
Example
AVER:COUN:AUTO OFF
Reset Condition
On reset, automatic averaging is enabled.
Query
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO?
The query enters a 1 or 0 into the output buffer indicating whether
automatic filter length is enabled or disabled.
Query Example
AVER:COUN:AUTO?
EPM-P Series Power Meters Programming Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
[SENSe[1]]|SENSe2:AVERage:COUNt:AUTO <boolean>
:
:COUN
AVER
1 is returned when automatic filter length is enabled.
0 is returned when automatic filter length is disabled.
SENSe Subsystem
Space
:AUTO
0|OFF
1|ON
ONCE
?
This command disables automatic
filter length selection for
channel A.
This command queries whether
automatic filter length selection is
on or off for channel A.