Site Preparation

Introduction

This section discusses the site preparation for the HP 5890 SERIES II
(hereafter referred to as HP 5890). The following topics are discussed:
C
C
C
You will need various supplies to prepare an area for the HP 5890.
Hewlett•Packard' s analytical supplies catalog describes the supplies
available from Hewlett•Packard.You can obtain a copy from your local
sales office.

Environmental considerations

This instrument operates in the 0
5% to 95% relative humidity range. We recommend an environment
comfortable for the operators (reasonably constant temperature and
humidity) for optimum performance and instrument lifetime.
Corrosive substances, whether gas, liquid, or solid, may affect materials
and components used in the HP 5890 and should be avoided.

Space considerations

The HP 5890 requires a clear surface about 56 cm (22 in.) deep by 71 cm
(28 in.) wide, capable of supporting at least 41 kg (90 lbs). Another 25 cm
(10 in.) must be allowed behind the instrument to dissipate oven exhaust
air (up to 400
6
Environmental considerations
Power considerations
Gas considerations
^
C).
^
^
to 55
C temperature range and the