Environmental Requirements

Environmental Requirements
Ensure that the following environmental requirements are met before using this equipment.

Operation environment

Use the equipment in the following environment:
NOTE
Do not use the equipment in environments where condensation may occur.
For the environmental requirements of the equipment's storage, see "Environment
conditions" on page 275.

Ventilation space at installation site

To ensure the specifications and measurement accuracy of the product, you must keep
ambient temperature around the product within the specified range by providing
appropriate cooling clearance around the product or, for the rackmount type, by forcefully
air-cooling inside the rack housing. For information on ambient temperature to satisfy the
specifications and measurement accuracy of the product, refer to "Measurement accuracy"
on page 255 in Chapter 12, "Specifications and Supplemental Performance
Characteristics."
When the ambient temperature around the product is kept within the temperature range of
the operating environment specification (refer to "Operating Condition" on page 275), the
product conforms to the requirements of the safety standard. Furthermore, under that
temperature environment, it has been confirmed that the product still conforms to the
requirements of the safety standard when it is enclosed with cooling clearance as follows:
20
5 °C to 40 °C
Temperature
Relative
20 % to 80 %, wet bulb temperature ≤ 29 °C (disk drive
humidity
condition)
15 % to 90 %, wet bulb temperature ≤ 29 °C (disk drive
non-operating condition)
Altitude
0 to 2,000 m (0 to 6,561 feet)
*1. Built-in floppy disk
≥ 180 mm
Back
≥ 60 mm (both right and left)