Using Device Clear To Halt Measurements - Keysight 34420A User Manual

Nano volt/micro ohm meter
Table of Contents

Advertisement

Using Device Clear to Halt Measurements

Device clear is an IEEE-488 low-level bus message which can be used to halt
measurements in progress. Different programming languages and IEEE-488
interface cards provide access to this capability through their own unique
commands. The status registers, the error queue, and all configuration states are
left unchanged when a device clear message is received. Device clear performs
the following actions.
– All measurements in progress are aborted.
– The meter returns to the trigger "idle state".
– The meter's input and output buffers are cleared.
– The meter is prepared to accept a new command string.
– For RS-232 operation, sending the control-C character will perform the
Keysight 34420A User's Guide
equivalent operations of the IEEE-488 device clear message. The meter's DTR
(data terminal ready) handshake line will be true following a device clear
message. See
"DTR / DSR Handshake Protocol"
details.
Remote Interface Reference
on page 194 for further
4
203

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents