Using Device Clear

Device Clear is an IEEE-488 low-level bus message that you can use to return the instrument to a respons-
ive state. Different programming languages and IEEE-488 interface cards provide access to this capability
through their own unique commands. The status registers, the error queue, and all configuration states
are left unchanged when a Device Clear message is received.
Device Clear performs the following actions:
If a measurement is in progress, it is aborted.
l
The instrument returns to the trigger "idle" state.
l
The instrument's input and output buffers are cleared.
l
The instrument is prepared to accept a new command string.
l
An overlapped command, if any, is terminated with no "Operation Complete" indication (applies to
l
INIT).
The ABORt command is the recommended method to terminate a measurement.

Commands by Subsystem

ABORt
CALCulate:CLEar[:IMMediate]
CALCulate:AVERage Subsystem
CALCulate:LIMit Subsystem
CALCulate:SCALe Subsystem
CALCulate:SMOothing Subsystem
CALCulate:TCHart[:STATe]
CALCulate:TRANsform:HISTogram Subsystem
CALibration Subsystem
CONFigure Subsystem
DATA Subsystem
DISPlay Subsystem
FETCh?
FORMat Subsystem
HCOPy Subsystem
IEEE 488.2 Common Commands
INITiate[:IMMediate]
LXI Subsystem
Keysight True volt Series Operating and Service Guide
SCPI Programming Reference