Insert artwork here.
Initial Observations - Normal Power-up Sequence
Prepare the analyzer for observation by performing these steps:
1. Disconnect all accessories and peripherals.
2. Remove the disk (if any) from the internal disk drive.
3. Connect a known-good cable between the RF OU and RF IN ports.
Switch ON the analyzer and watch for the proper power-up sequence:
1. Immediately after the analyzer's line power is switched on, the disk drive light activates
brie
y and a \beep" is audible from the CPU.
2. he CR lists the bootROM self-tests being performed. ( he CR may take a few seconds to
warm up.) If all bootROM self-tests pass, the analyzer beeps twice.
3. he disk drive light illuminates brie
y and the rmware loads from ROM.
4. he analyzer runs all of the remaining self-tests and the CR displays the PASS/FAIL status
of the self-tests.
5. During the next few seconds, these messages appear on CR :
a.
Initializing
b.
Calculating correction coefficients
c.
Initializing detector corrections
d. Other messages may appear brie
y, depending on the analyzer's state at the time that the
power was switched o.
Figure 4-2. Connecting the Extender Board
T roubleshoot ng and Block D agrams
4-5