Instrument Functions: M - O
Marker Fctn
3.2.4 Function Off
Turns off marker functions (
Delta markers will remain on screen.
NOTE
Key Path:
Marker Fctn
Remote Command:
See
"Marker Fctn" on page 179
Example:
CALC:MARK2:FUNC OFF turns the marker 2 function off.
3.2.5 Marker Count
Accesses the marker count menu.
Key Path:
Marker Fctn
3.2.5.1 Marker Count
Turns the marker frequency counter on and off for any active marker. If no marker is
active before
Marker Count
asterisk (*) may appear in the upper-right area of the display along with the message Cntr
1 (the number in the message depends on the active marker). If the marker count function
is on and you change the active marker, the new active marker will use marker count. If
the frequency counter function is on with only one active marker and that marker is
turned off, then the frequency counter function is turned off. If the frequency counter
function is on with multiple markers turned on and only one is turned off, the frequency
counter function stays on.
frequency offset function.
In
the counter continues to function, counting any signal near the center
Zero Span
frequency of the analyzer.
Setting
NOTE
Marker Fctn
Key Path:
Marker Fctn
State Saved:
If
Marker Count
Factory Preset: Off
182
and
Band/Intvl Power
for the command to select the function.
is pressed, a marker is activated at the center of the display. An
frequency readings are not affected by the
Marker Count
to Off does not turn
,
Marker Count
is on, that setting is saved in the instrument state.
).
Marker Noise
off.
Marker Count
Chapter 3