ESA Functional Tests

Marker Count Accuracy

Marker Count Accuracy

Procedure

1. Press
,
System
Power On/Preset
for the preset routine to finish. Set the analyzer to measure the marker count accuracy
by pressing the following keys:
,
Frequency
Center Freq
, 10,
SPAN
MHz
,
, 100,
BW/Avg
Res BW
,
Freq Count
Resolution
2. Press
, then wait for a count to be taken (it may take several seconds).
Peak Search
3. Record the Cntr1 frequency as the Marker Frequency Readout in
Table 4-3
Frequency Readout and Marker Count Accuracy Worksheet
Span
10 MHz
1.48988 GHz
100 kHz
1.4899988 GHz
Marker Count
1.489999999 GHz
Accuracy w/Counter
Resolution at 1 Hz
166
,
(Factory),
Preset Type
, 1.490,
GHz
kHz
, 1,
Hz
Marker
Minimum
Frequency
Readout
on the analyzer, then wait
Preset
Table
4-3.
Maximum
1.49012 GHz
1.4900012 GHz
1.490000001 GHz