Freq Count

Activates the
turns the marker count function on.
Marker Count
Marker Count (On)
before
On Off
of the display. Press
Press
uncoupled value.
Marker Count frequency readings are not affected by the frequency
NOTE
offset function.
An asterisk (*) may appear in the upper-right area of the display along
with the message Cntr 1 (the number in the message depends on the
active marker). The ratio of the resolution bandwidth to span must be
greater than 0.002 for the marker count function to work properly.
Marker Count: Widen Res BW appears on the display if the bandwidth
to span ratio is less than 0.002. Widen RES BW indicates that the
resolution bandwidth must be increased or the span decreased.
Key Access:
Allows the resolution of the marker counter to be selected manually or
Resolution
auto-coupled. The marker counter has a resolution range of 1 Hz to
Auto Man
100 kHz. The available resolution values are 1 Hz, 10 Hz, 100 Hz,
1 kHz, 10 kHz, and 100 kHz. The resolution can be changed by using
the step keys or by entering the resolution using the numeric keypad or
knob. The marker counter resolution can be auto coupled to the span by
pressing
Key Access:
Chapter 2
Marker Normal
turns on the marker counter. If no marker is active
is pressed, a marker is activated at the center
Marker Count (On)
Marker Count (Off)
to change the marker counter resolution to an
Resolution (Man)
Freq Count
.
Resolution (Auto)
Freq Count
function if there are no markers, and then
to turn the marker counter off.
Front-Panel Key Reference
Freq Count
97