[:Sense]:Average:count <Integer; [:Sense]:Average:count - Agilent Technologies N1996A-503 User Reference

Csa spectrum analyzer
Table of Contents

Advertisement

running – the trace is not updated until the last average count has been
reached.
Example
Key Path
Repeat Max Hold Sets
Avg Mode
captures and holds the highest magnitude data at the end of each sweep count.
NOTE
With large average counts and/or slow traces, the trace from the last
completed result remains on the display while
running – the trace is not updated until the last average count has been
reached. (Maximum data points are held with each consecutive sweep)
Example
Key Path
Avg Number
Sets the average count number (the number of sweeps) used with measurement averaging.
Mode
Remote Command
Example
Preset
Min
Max
Key Path
Avg Type
Accesses averaging settings that determine how the signal will be measured. You can set
Chapter 3
AVER:TCON REP
Meas Setup, Avg Mode, Repeat
to repeat with max hold function. Repeat with max hold
AVER:TCON RMAX
Meas Setup, Avg Mode, Repeat Max Hold
SA
[:SENSe]:AVERage:COUNt <integer>

[:SENSe]:AVERage:COUNt?

AVER:COUN 50
100
1
4096
Meas Setup, Avg Number
Meas Off (Spectrum Analyzer) Measurement
Repeat
Max Hold is
Mode
59