Caution
Do not exceed the operating input power, voltage, and current
level and signal type appropriate for the instrument being used, refer to
your instrument's Function Reference.
Electrostatic discharge(ESD) can damage the highly sensitive
microcircuits in your instrument. ESD damage is most likely to occur as
the test fixtures are being connected or disconnected. Protect them from
ESD damage by wearing a grounding strap that provides a high
resistance path to ground. Alternatively, ground yourself to discharge any
static charge built-up by touching the outer shell of any grounded
instrument chassis before touching the test port connectors.

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies E5071C ENA Series

  • Page 1 Caution Do not exceed the operating input power, voltage, and current level and signal type appropriate for the instrument being used, refer to your instrument's Function Reference. Electrostatic discharge(ESD) can damage the highly sensitive microcircuits in your instrument. ESD damage is most likely to occur as the test fixtures are being connected or disconnected.
  • Page 2 When you notice any of the unusual conditions listed below, immediately terminate operation and disconnect the power cable. Contact your local Agilent Technologies sales representative or authorized service company for repair of the instrument. If you continue to operate without repairing the instrument, there is a potential fire or shock hazard for the operator.
  • Page 3 Herstellerbescheinigung GER  AUSCHEMISSION < 70 dB am Arbeitsplatz normaler Betrieb nach DIN 45635 T. 19 Manufacturer's Declaration ACOUSTIC NOISE EMISSION < 70 dB operator position normal operation per ISO 7779...
  • Page 4 This product complies with the essential requirements of the following applicable European Directives, and carries the CE marking accordingly: The Low Voltage Directive 73/23/EEC, amended by 93/68/EEC The EMC Directive 89/336/EEC, amended by 93/68/EEC To obtain Declaration of Conformity, please contact your local Agilent Technologies sales office, agent or distributor.
  • Page 5 ・ This equipment is tested with stand-alone condition or with the combination with the accessories supplied by Agilent Technologies against the requirement of the standards described in the Declaration of Conformity. If it is used as a system component, compliance of related regulations and safety requirements...
  • Page 7 Agilent E5071C ENA Series RF Network Analyzers Installation Guide Fifth Edition Manufacturing No. E5071-90040 August 2008...
  • Page 8 This document contains proprietary information that is protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of Agilent Technologies. Microsoft®,MS-DOS®,Windows®,Visual C++®,Visual Basic®,VBA® and Excel® are registered trademarks of Microsoft Corporation.
  • Page 9 To avoid the danger of introducing additional hazards, do not install substitute parts or perform unauthorized modifications to the instrument. Return the instrument to an Agilent Technologies Sales and Service Office for service and repair to ensure that safety features are maintained in operational condition.
  • Page 10 Safety Symbols General definitions of safety symbols used on the instrument or in manuals are listed below. Instruction Manual symbol: the product is marked with this symbol when it is necessary for the user to refer to the instrument manual. Alternating current.
  • Page 11 Certification Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institution’s calibration facility or by the calibration facilities of other International Standards Organization members.
  • Page 12 Typeface Conventions Sample (bold) Boldface type is used when a term is defined or emphasized. Sample (Italic) Italic type is used for emphasis and for titles of manuals and other publications. menu/button/box/tab Indicates a menu/button/box/tab on the screen Sample labeled “Sample” which can be selected/executed by clicking.
  • Page 13: Table Of Contents

    Contents 1. Introduction Manuals for E5071C ..............10 Contents of This Manual.
  • Page 14 Contents...
  • Page 15: Introduction

    Introduction This chapter introduces the documentation for Agilent E5071C and the overview of this manual to use this manual efficiently. See this manual first when you use the E5071C for the first time.
  • Page 16: Manuals For E5071C

    Introduction Manuals for E5071C Manuals for E5071C Agilent provides the following three manuals for E5071C. The latest version of all docu- mentations can be downloaded from http://ena.tm.agilent.com/e5071c/manuals/. Installation Guide The installaton guide (this manual) provides start up setup information when you use the E5071C for the first time and troubleshooting information when the Windows cannot be boot up.
  • Page 17: Installation

    Installation This chapter provides information about how to set up the Agilent E5071C.
  • Page 18 Installation Contents of this Chapter Checking the Shipment on page 13 After you receive the analyzer, check all the items in the packing container. Environmental Requirements on page 14 Describes the system requirements needed to install the E5071C and how to secure space for heat radiation.
  • Page 19: Checking The Shipment

    Step 3. By referring to the furnished contents list, check that all packaged items supplied with the analyzer have been provided as per the specified options. Step 4. After checking, if one of the following applies, contact your nearest Agilent Technologies sales and service office.
  • Page 20: Environmental Requirements

    Installation Environmental Requirements Environmental Requirements Set up the E5071C where the following environmental requirements are met. Operating Environments Ensure that the operating environment meets the following requirements. 5 °C to 40 °C Temperature 23 °C ± 5 °C (< 1 °C deviation from the temperature Temperature range at the error-correction when performing the error-correction)
  • Page 21: Ventilation Requirements

    Installation Environmental Requirements Ventilation Requirements To ensure safety requirements, the specifications and measurement accuracy of the analyzer, you must keep environmental temperature within the specified range by providing appropriate cooling clearance around the analyzer or, for the rackmount type, by forcefully air-cooling inside the rack housing.
  • Page 22: Protection Against Electrostatic Discharge (Esd)

    Installation Environmental Requirements Protection Against Electrostatic Discharge (ESD) Set up a static-safe work-station to protect the electronic components from the damage by the electrostatic discharge (ESD) as shown in Figure 2-2. Figure 2-2 Example of the static-safe work station Ensuring Adequate Free Space around Analyzer for Immediate Disconnection of Power Cable in Case of Emergency As described in “Disconnection from Supply Source”...
  • Page 23: Installing Front Handles/Rack Mounting Flanges

    Installation Installing Front Handles/Rack Mounting Flanges Installing Front Handles/Rack Mounting Flanges The E5071C can be installed on a workbench or in a rack. This section describes how to install the front handles (Option 1CN) used for moving or transporting the instrument, and how to install the analyzer in an equipment rack as part of a measurement system (Option 1CM: without the handles, Option 1CP: with the handles).
  • Page 24: How To Install The Handle Kit (Option 1Cn)

    Installation Installing Front Handles/Rack Mounting Flanges Figure 2-3 Installing front handle/rack-mount kits How to Install the Handle Kit (Option 1CN) The handle kit is used for transport and relocation of the E5071C. While referring to Figure 2-3, install the handle kit by following these steps. Step 1.
  • Page 25: How To Install The Rack-Mount Kit (Option 1Cm)

    Installation Installing Front Handles/Rack Mounting Flanges How to Install the Rack-mount Kit (Option 1CM) The rack-mount kit includes two flanges (locking side plates) for mounting the E5071C on a rack (482.6 mm/19 inches) conforming to the EIA Standard. While referring to Figure 2-3, install the rack-mount kit by following these steps.
  • Page 26: Connecting The Accessaries

    Installation Connecting the Accessaries Connecting the Accessaries You need to connect some accessaries before use. Connecting the BNC Adaptor (Option 1E5) When the E5071C is equipped with Option 1E5 (high-stability frequency standard), connect the BNC adaptor from the Ref Oven terminal to the Ref In terminal on the rear panel as shown in Figure 2-4.
  • Page 27: Power Supply

    Installation Power Supply Power Supply Before turning on the E5071C power, check the following. Verification of the Power Supply Confirm that the power supplied to the E5071C meets the following requirements: Requirements Voltage 90 to 132 VAC or 198 to 264 VAC Frequency 47 to 63 Hz Maximum power consumption...
  • Page 28: Blown Fuses

    Blown Fuses If the fuse appears to have blown during operation, this instrument may be subject to fail- ure and must be reapired. For any assistance, contact Agilent Technologies Customer con- tact centre listed at the end of this guide.
  • Page 29: Starting The E5071C

    Installation Starting the E5071C Starting the E5071C This section describes how to turn on/off the E5071C power and how to cut off the power supply in an emergency. Turning the Power ON and OFF Perform the following steps to turn the power ON or OFF. To Check the Removable Hard Disk Drive Procedure When the power is turned on, it is required to confirm the notch geometry is set as follows.
  • Page 30 Installation Starting the E5071C This operation turns ON the power, and the E5071C starts the self-test. Step 4. Confirm that the self-test indicates normal operation. Normal operation is confirmed by the self-test if no error message appears. Turning the Power OFF Step 1.
  • Page 31: Disconnection From Supply Source

    Installation Starting the E5071C Disconnection from Supply Source The power supply of the E5071C is cut off by disconnecting the plug of the power cable (on either AC outlet side or E5071C side). When it is necessary to disconnect the power supply in order to avoid shock hazards, etc., pull out the power cable plug from either the AC outlet side or the E5071C side.
  • Page 32: Removing / Mounting Removable Hard Disk (For Option 017 User)

    Installation Removing / Mounting Removable Hard Disk (For Option 017 User) Removing / Mounting Removable Hard Disk (For Option 017 User) For details of removing / mounting removable hard disk, refer to the following website. http://ena.tm.agilent.com/e5071c/manuals/webhelp/eng/ Chapter 2...
  • Page 33: Initial Registration Of E5071C

    Installation Initial Registration of E5071C Initial Registration of E5071C When you start up the E5071C at the first time, you need to perform the initial registration of the Windows XP operating system of the E5071C. NOTE You cannot use the front panel keys during the initial registration of the E5071C therefore connect the mouse and the keyboard before turning on the power.
  • Page 34 Installation Initial Registration of E5071C Step 3. In the Windows XP Professional Setup dialog box, select the I accept this agreement and click the button (Figure 2-9). Next > Figure 2-9 Windows XP Professional Setup dialog box (1/2) e5071cse1074 Step 4. In the next dialog box , input in the box.
  • Page 35: Calibration Of The Touch Screen

    Installation Calibration of the Touch Screen Calibration of the Touch Screen After E5071C measurement screen appears, you have to calibrate the touch screen. Follow the procedure described below to calibrate the touch screen. Step 1. Press Step 2. Press Service Menu Step 3.
  • Page 36 Installation Calibration of the Touch Screen Chapter 2...
  • Page 37: Troubleshooting

    Troubleshooting This chapter describes the troubleshooting during start up and the procedure of the operating system (OS) recovery when the Windows XP OS has been damaged.
  • Page 38: Troubleshooting During Startup

    Troubleshooting Troubleshooting during Startup Troubleshooting during Startup When you encounter the problem during start up, see Table 3-1. System recovery saves most of problems. Table 3-1 Troubleshooting during startup Symptom Solution Turning on (|) the standby switch does not Confirm that the power cable is properly start up the system.
  • Page 39: System Recovery

    Troubleshooting System Recovery System Recovery By executing system recovery, you can return the system of the E5071C (the Windows operating system and the firmware) to the factory state (at the time of purchase CAUTION Strictly follow the steps described below. If you do any operation other than the following steps, the system may not be recovered.
  • Page 40: Procedure To Execute The Factory Recovery

    Troubleshooting System Recovery Procedure to execute the factory recovery This section describes how to return the contents of the C drive to the factory state. NOTE You need a keyboard for this operation. Step 1. Shut down the E5071C. Step 2. Disconnect all of the USB device from the USB ports. Step 3.
  • Page 41 Troubleshooting System Recovery Step 6. Select Restore Figure 3-1 Activate Acronis Startup Recovery Manager e5071cse1079 Step 7. A confirmation dialog box is displayed. Push the enter to select to start the recovery of the factory backup. The recovery takes a few minutes depending on the amount of data. Figure 3-2 Recover Factory Backup e5071cse1034...
  • Page 42 Troubleshooting System Recovery CAUTION Never turn off the power during the system recovery because doing so may cause serious damage to the E5071C. Chapter 3...
  • Page 43: Procedure To Create The User Backup Image

    Troubleshooting System Recovery Procedure to create the user backup image This section describes how to create the user backup image. The C and D drives’ contents saved in this procedure are recalled when the user recovery function is executed. Step 1. Close the E5071C application by pressing >...
  • Page 44 Troubleshooting System Recovery Step 6. Click button. Next e5071cse1126 Step 7. Select “The entire disk contents or individual partition,” then click button. Next e5071cse1127 Chapter 3...
  • Page 45 Troubleshooting System Recovery Step 8. Select the C drive and D drive. e5071cse1128 Step 9. Select the destination folder and input a file name, then click button. It is Next recommended to save the file to a directory which has enough space like the E drive. Also, never save the contents of the drive to the C, D or F drive.
  • Page 46 Troubleshooting System Recovery Step 10. Select “Set the options manually” and click button. Next e5071cse1130 Step 11. Select “Backup priority” in the left field and “High” in the right field, then click Next button. e5071cse1131 Chapter 3...
  • Page 47 Troubleshooting System Recovery Step 12. If necessary, input comments and click button. The comments can be omitted. Next e5071cse1132 Step 13. Click button. Proceed e5071cse1133 Chapter 3...
  • Page 48 Troubleshooting System Recovery Step 14. The following screen appears during backup. e5071cse1134 Step 15. The following screen appears when the user backup image was created. Click e5071cse1135 CAUTION Never turn off the power during creating the user backup image because doing so may cause serious damage to the E5071C.
  • Page 49: Procedure To Execute The User Recovery Function

    Troubleshooting System Recovery Procedure to execute the user recovery function Returns the contents of the C and D drives to a user-specified state. To use this function, you must create the user backup image in advance. For more information, see the description “Procedure to create the user backup image”...
  • Page 50 Troubleshooting System Recovery Step 6. Click button. Next e5071cse1137 Step 7. Select the saved backup image file, then click button. Next e5071cse1138 Chapter 3...
  • Page 51 Troubleshooting System Recovery Step 8. Select “Restore disks or partitions,” then click button. Next e5071cse1139 Step 9. Select the C drive, then click button. Next e5071cse1143 Chapter 3...
  • Page 52 Troubleshooting System Recovery Step 10. Select the C drive to restore, then click button. Next e5071cse1144 Step 11. Select “Active” and click button. Next e5071cse1145 Chapter 3...
  • Page 53 Troubleshooting System Recovery Step 12. Click button. Next e5071cse1146 Step 13. Select “Yes, I want to assign a logical drive letter to the restored partition” and click Next button. e5071cse1147 Chapter 3...
  • Page 54 Troubleshooting System Recovery Step 14. Select “Yes, I want to restore another partition or hard disk drive,” then click button. Next e5071cse1148 Step 15. Select the D drive to restore, then click button. Next e5071cse1149 Chapter 3...
  • Page 55 Troubleshooting System Recovery Step 16. Select the D drive to restore, then click t button. e5071cse1150 Step 17. Select “Logical” and click button. Next e5071cse1151 Chapter 3...
  • Page 56 Troubleshooting System Recovery Step 18. Click button. Next e5071cse1152 Step 19. Select “Yes, I want to assign a logical drive letter to the restored partition,” then click Next button. e5071cse1153 Chapter 3...
  • Page 57 Troubleshooting System Recovery Step 20. Select “No, I do not.” and click button. Next e5071cse1154 Step 21. Select “Set the options manually” and click button. Next e5071cse1155 Step 22. Select “Restoration Priority” in the left field and “High” in the right field, then click Next Chapter 3...
  • Page 58 Troubleshooting System Recovery button. e5071cse1156 Step 23. Click button. Proceed e5071cse1157 Chapter 3...
  • Page 59 Troubleshooting System Recovery Step 24. Click button. Reboot e5071cse1158 CAUTION Never turn off the power during the recovery of the user backup image because doing so may cause serious damage to the E5071C. Step 25. After boot up, perform “Initial Registration of E5071C” on page 27 and “Calibration of the Touch Screen”...
  • Page 60 Troubleshooting System Recovery Chapter 3...
  • Page 61 REGIONAL SALES AND SUPPORT OFFICES For more information about Agilent Technologies test and measurement products, applications, services, and for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir. You can also contact one of the following centers and ask for a test and measurement sales representative.