2
Operation and Features
& SOUR:MEM:OUTP ON, (@1)
& SOUR:MEM:VOLT:MEAS, (@1)
& SOUR:MEM:LIST:STOR, (@1)
& SOUR:MEM:TRIG (@1)
N O T E
Memory list commands execution state
Figure 2- 15
when executing commands from the memory list using the external trigger
or remote trigger command, and the status operation condition register bit
that is involved. The WTG bit set indicates that the channel is armed and
waiting for trigger. The DTG bit set indicates that the channel is busy and
will ignore all incoming commands.
58
1 When the measurement results reaches 200, the next result will wrap back and
replace the first result at the buffer.
2 When the last executing command in the memory list is to change the source
range, this command will not take effect at the output until subsequent source
command is executed.
3 When a command is stored into the memory list, the active voltage and current
range at that time is stored as well. If a command in the memory list is
executed, these ranges will be restored to the active ranges.
and
Figure 2- 16
shows the states that a channel goes through
This command stores the "set output on"
command into the active memory list of
channel 1.
This command stores the "measure voltage
command" into the active memory list of
channel 1.
This command stores all commands from the
active memory list of channel 1 into the
nonvolatile memory. This command can be
omitted if this operation is not required.
This command executes all the commands from
the active memory list of channel 1.
U2722A/U2723A User's Guide