Trigger[:Sequence]:Delay <Time; Trigger[:Sequence]:Delay; Trigger[:Sequence]:External:slope Positive|Negative; Trigger[:Sequence]:External:slope - Agilent Technologies N1996A-503 User Reference

Csa spectrum analyzer
Table of Contents

Advertisement

Mode
Meas Off (Spectrum Analyzer) Measurement
Dependencies / Couplings
Key Path
Trig Slope Controls the trigger polarity. It is set positive to trigger on a rising edge and
negative to trigger on a falling edge.
Remote Command
Example
Preset
State Saved
Dependencies / Couplings
Key Path
Trig Delay Controls a time delay during which the analyzer will wait to begin a sweep
after meeting the trigger criteria. You can use negative delay to pre-trigger the instrument
in time domain, but not in swept spans.
Remote Command
Example
Preset
State Saved
Min
Max
Dependencies / Couplings
74
The range of the Video Trigger Level is dependent on the Reference
Level.
Trigger level selection is not available with free run or external
triggering.
Meas Setup, Trigger, Trigger Level

:TRIGger[:SEQuence]:EXTernal:SLOPe POSitive|NEGative

:TRIGger[:SEQuence]:EXTernal:SLOPe?

TRIG:EXT:SLOP NEG
POSitive
Saved in State
Trigger slope selection is only available with external triggering.
Meas Setup, Trigger, Trig Slope
:TRIGger[:SEQuence]:DELay <time>

:TRIGger[:SEQuence]:DELay?

TRIG:DEL 100 ms
1 µs
Saved in State
-204.8 s
Note this is RBW dependent: -102400 / RBW (Hz) with a maximum
of -27 ms
+4.5 hours
Note this is RBW dependent: -102400 / RBW (Hz) with a minimum
of +2 s
Trigger delay selection is not available with free run triggering.