General Information
Required Equipment

Power Meter Accuracy Test

This test is intended for power meters used in testing the E5061A/E5062A. The "Dynamic
Accuracy Test" requires the use of a power meter that has been calibrated to a higher
accuracy than the standard power meter.
Power meters with options G12 and H12 specify an improved instrument accuracy over a
limited power range. (These power meters do not contain unique hardware.) A power
meter may be returned to the factory to have one of these options added to an existing
power meter or to renew the calibration for one of these options.
This test procedure is an alternative to returning the power meter to the factory. When a
power meter passes this test, it is considered to be calibrated for the G12 or H12 option
even though it has not been returned to the factory.
Power Meters That Can Be Tested Using This Procedure
This procedure assume that the E4419B power meter is being tested.
NOTE
It is recommended that the revision number for the power meter "Main Firmware" be
Ax.03.00 or higher.
Equipment Used for the Power Meter Accuracy Test
Equipment Type
Range calibrator
Precision digital
voltmeter
Power sensor
cable
Cable, BNC, 50 
Adapter for
connecting BNC
cable to DVM
inputs
Description of the Test
The power meter accuracy is verified for various power inputs and the actual readings are
recorded in a test record. A range calibrator is used to provide the reference inputs.
NOTE
It is recommended that a copy of the test record on page 24 be made, and the values be
recorded on the copy, thus preserving the original for future use.
20
Recommended Model
or Part Number
Agilent 11683A
Agilent 3458A
Agilent 11730A, p/n
8120-8319
Any
Any
*1.Required accuracy and resolution at the following voltage levels:
14 mV input: 0.0100% accuracy, 10 nV resolution
0.14 V input: 0.0050% accuracy, 100 nV resolution
0.45 V input: 0.0030% accuracy, 100 nV resolution
Alternate
Model or Part Number
None
Any with the required
*1
accuracy and resolution
Any equivalent
Any
Any