Command *Ese - Agilent Technologies HP E5574A User Manual

Optical loss analyzer
Table of Contents

Advertisement

syntax
description
parameter
example
command
syntax
description
response
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
HP-IB Programming
IEEE Common Commands
*ESE<wsp><value>
The Event Status Enable command *ESE sets bits in the Event
Status Enable register that enable the corresponding bit in the Event
Status Register ESR.
The register is cleared at power-on. The *RST and *CLS
commands do not affect the register.
The bit value for the register: 0 ≤ value ≤ 255.
Power On
6
User Request
5
Command Error
4
Execution Error
3
Device Dependent Error
2
Query Error
1
Request Control
Operation Complete
*ESE?
*ESE?
The Event Status Enable query *ESE? returns the contents of the
Event Status Enable register.
The bit value for the register: 0 ≤ value ≤ 255. See *ESE for
information on this register.
*ESE
110
128
64
32
16
8
4
2
1

Advertisement

Table of Contents
loading

Table of Contents