Ese - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Parameter
Query response
Semantics

*ESE

This command sets the bits of the Standard Event Status "Enable" Register. 1
enables, 0 masks.
*ESE mask_number
Parameter
mask_number
decimal or
non-decimal
numeric
mask_number <newline><^END>
mask_number is decimal numeric value that is the sum of the binary-weighted
values for each bits. Response data type is NR1.
The Standard Event Status "Enable" Register determines which bits of the Standard
Event Status Register are enabled. Enabled bits are ORed together, and the result is
reported to bit5 of the Status Byte Register.
The Standard Event Status "Enable" Register consists of 16 bits, but only the lower
8 bits are used, which correspond to the bits of the Standard Event Status Register.
Refer to Chapter 3 for details about the Standard Event Status "Enable" Register
operation.
You can send a mask_number parameter in decimal numeric or non-decimal
numeric type.
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Type
numeric value that is the sum of the
binary-weighted values for the desired
bits
Common Commands
*ESE
Explanation
4-5