Power Meter Operation
Zeroing and Calibrating the Power Meter

Zeroing and Calibrating Using TTL Inputs

You can use the TTL inputs on the rear panel Rmt I/O port to initiate zero and
calibration cycles on the power meter. The connector is an RJ-45 series shielded
modular jack with the TTL input pins connected as shown in Figure 2-3.
Figure 2-3: Rmt I/O Port TTL Inputs
1 2 3 4 5 6 7 8
The TTL inputs are active low and control the zero and calibration functions as shown
in Table 2-2
Table 2-2: TTL Inputs Control Logic
Input 1
2-12
Pin Number
1
2
3
4
5
6
7
8
Input 2
1
1
1
0
0
1
0
0
Connection
none
Ground
Upper Window TTL Output
Lower Window TTL Output
TTL Input 1
TTL Input 2
Ground
Ground
Operation
None
CAL
ZERO
CAL
Agilent E4418B User's Guide