Power Meter Operation
Setting Measurement Limits
default values depends on the measurement units in the currently
selected window - see Table 2-6.
Table 2-6: Range of Values for Window Limits
Window
Max
Units
dB
+200 dB
dBm
+230 dBm
%
999.9 X%
W
100.000 XW
The window based limits can also be set to output a TTL logic level at the
rear panel Rmt I/O port when the predefined limits are exceeded. You can
switch the rear panel TTL outputs on or off; set the TTL output level to
active high or low; and determine whether the TTL output represents an
over limit condition, under limit condition or both.
The TTL connector is an RJ-45 series shielded modular jack assembly
with the TTL output pins connected as shown in Figure 2-14.
2-48
Min
Max
-180 dB
60 dB
-150 dBm
90 dBm
100.0 a%
100.0 M%
1.000 aW
1.000 MW
Agilent E4419B User's Guide
Default
Min
-120 dB
-90 dBm
100.0 p%