[:Sense] - Agilent Technologies ESA Series User's/Programmer's Reference

Core spectrum analyzer functions
Hide thumbs Also See for ESA Series:
Table of Contents

Advertisement

Remote Command Reference
SENSe Subsystem
SENSe Subsystem
Sets the instrument state parameters so that you can measure the input signal.
SENSe subsystem commands used for measurements in the
menus are located in Volume 2. These commands may be used only to set
Setup
parameters of a specific measurement when the measurement is active.
Auto-range Dwell Time
[:SENSe]:ARDTime <time>
[:SENSe]:ARDTime?
Sets the auto-range dwell time for the EMI Detectors. The amplitude
ranging process optimizes the available dynamic range for the
measurement. The dwell time is the time spent auto-ranging. You may
need to set this value if you have a signal with a very low PRF signal
(< 5Hz). Dwell time must be > 1/PRF. (See also
SENSe:DETector:FUNCtion:EMI )
Factory Preset
and *RST:
0.2 s
Range:
0.001 to 2000 s (The actual minimum value varies. It is
dependent on the number of trace points and the ADC period.)
Remarks:
This command is only available with Option AYQ installed.
History:
Added with firmware revision A.14.00.
Front Panel
Access:
Det/Demod
304
,
EMI Detector
and
MEASURE
Meas
Chapter 5