[:Sense]:Acpower:sweep:time - Agilent Technologies N1996A-503 User Reference

Csa spectrum analyzer
Table of Contents

Advertisement

Example
Preset
State Saved
Range
Key Path
Control/Sweep
Accesses the sweep and measurement control settings that can be changed when using the
ACP measurement.
Mode:
Key Path
Restart, Single, Cont
The sweep control functionality is shared across all the Modes. See
Functions" on page 252
Sweep Time Query (Remote Command Only)
The sweep time value can be queried using the [:SENSe]:ACPower:SWEep:TIME?
command. The ACP measurement is a frequency domain measurement only, so the Sweep
Time key is disabled (not available) in the ACP measurement.
Mode
Remote Command
Example
Notes
Preset
Key Path
Chapter 3
:DISP:ACP:WIND:TRACe:Y:PDIV 10dB
10.00 dB / Div
Saved in State
1.0 to 20
AMPTD Y Scale, Scale/Div
SA
Front-panel key
for details about how they operate.
SA

[:SENSe]:ACPower:SWEep:TIME?

ACP:SWE:TIME?
In ACP measurement (Frequency Domain measurement), you
can only query the "sweep" time. Trying to set the Sweep Time
will generate a system error message
The default value is auto-calculated and will depend on the
Frequency range of the Analyzer.
N/A - Remote Feature Only
Adjacent Channel Power—ACP (I&M) Measurement
"Control/Sweep