Compatible SCPI Commands
Supported SCPI Commands
State Saved
Min
Max
STATus Subsystem
The STATus subsystem provides the status reporting system commands. The status registers are not
influenced by *RST.
STATus:OPERation[:EVENt]?
Queries the contents of the EVENt section in the STATus:OPERation register. After the readout, the
contents of the EVENt section are deleted.
Remote Command
Example
Remote Command Notes
Preset
STATus:OPERation:CONDition?
Queries the CONDition section in the STATus:OPERation register. The current hardware status is
reflected in the value returned. The contents of the CONDition section are not deleted after the readout.
Remote Command
Example
Remote Command Notes
Preset
STATus:OPERation:ENABle
Specifies the bits of the ENABle section in the STATus:OPERation register.
This mode supports the maximum value of 32767, not 65535 as in FSx.
Remote Command
Example
Remote Command Notes
Preset
Min
Max
152
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html
Saved in instrument state.
1
8001
:STATus:OPERation[:EVENt]?
STAT:OPER?
FSE, FSP, FSU
0
:STATus:OPERation:CONDition
STAT:OPER:COND?
FSE, FSP, FSU
0
:STATus:OPERation:ENABle <integer>
:STATus:OPERation:ENABle?
STAT:OPER:ENAB 1
FSE, FSP, FSU
0
0
32767