Status:operation Condition And Event Enable Registers - Agilent Technologies ESA Series User's/Programmer's Reference

Core spectrum analyzer functions
Hide thumbs Also See for ESA Series:
Table of Contents

Advertisement

Bit

STATus:OPERation Condition and Event Enable Registers

The STATus:OPERation condition register continuously monitors the hardware and firmware status of
the analyzer, and is read-only. To query the register, send the :STATus:OPERation:CONDition?
command. The response will be the decimal sum of the bits that are set to 1. For example, if bit number 9
and bit number 3 are set to 1, the decimal sum of the 2 bits is 512 plus 8. So the decimal value 520 is
Chapter 4
Use Status Registers to Determine the State of Analyzer Events and Conditions
Decimal
Description
Value
1
2
Reserved: This bit is not used by the analyzer, but is for
future use with other Agilent products.
2
4
Reserved: This bit is not used by the analyzer, but is for
future use with other Agilent products.
3
8
Sweeping: A 1 in this bit position indicates that a sweep is
in progress.
a
16
Measuring: A 1 in this bit position indicates that a
4
measurement is in progress.
a
32
Waiting for Trigger: A 1 in this bit position indicates
5
that a measurement is in a "wait for trigger" state.
6
64
Reserved: This bit is not used by the analyzer, but is for
future use with other Agilent products.
7
128
Reserved: This bit is not used by the analyzer, but is for
future use with other Agilent products.
b
256
Paused: A 1 in this bit position indicates that the instrument
8
is in the paused state of the measurement.
9
512
Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
10
1024
Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
11
2048
Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
12
4096
Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
13
8192
Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
14
16384
Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
15
32768
Always Zero (0)
a. The description of this bit refers to any measurement under the
b. This bit applies to ESA optional measurement personalities only, and may or may not
be implemented in all such personalities.
Status Registers
MEASURE
key.
211