Saving The Calibration Factor; Connecting Thru Standard - Agilent Technologies ENA Series Installation And Quick Start Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Quick Start Guide
Measurement Example of a Bandpass Filter
Step 5. Connect the THRU standard betwween the measurement cables as shown in Figure 3-14,
and measure the thru calibration data. After measuring the thru calibration data, symbol √
is displayed to the left of the
Figure 3-14
Connecting the THRU standard
Step 6. Make the full 2-port calibration measurement DONE. The calibration factor is calculated
based on the calibration data acquired, and the error correction is turned ON.
Step 7. Select the type in which the data is to be saved before saving the calibration factor
(calculated based on the calibration data).
Step 8. Store the calibration file to the disk of the E5061A/E5062A. The symbol "X" appearing in
the operations below represent the assigned numbers to be used when the file is saved.
For more information on saving/reading internal data, see Chapter "Data Output" in the
E5061A/E5062A User's Guide.
54
Port 1-2 Thru
[Cal] - Calibration - 2-Port Cal - Transmission - Port 1-2 Thru
[Cal] - Calibration - 2-Port Cal - Done
[Save/Recall] - Save Type - State & Cal
[Save/Recall] - Save State - State 0X
button.
Chapter 3

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5061aE5062aE5052a

Table of Contents