How To Set Measuring Conditions; Presetting The 4287A; How To Set The Measurement Point (Frequency/Averaging Factor/Oscillator Level) - Agilent Technologies Twelfth 4287A RF LCR Operation Manual

Table of Contents

Advertisement

Learning Basic Operations

How to Set Measuring Conditions

How to Set Measuring Conditions
This sections explains how to set the measuring conditions.
NOTE
The following explanation of how to set measuring conditions assumes that you are using
the keys on the front panel of the 4287A. If you want to know the setting procedure when
using a keyboard or mouse, see "4287A Key Operation Overview" on page 99.

Presetting the 4287A

Press the
initially set state called "Preset state."
How to set the measurement point (frequency/averaging
factor/oscillator level)
With the 4287A, you must enter the desired measurement point along with its conditions in
the measurement point setup display before starting measurement. The following
procedure shows a case when four measurement points with different frequencies (i.e., 1
MHz, 10 MHz, 100 MHz, and 1 GHz) are set up.
For all four points, the averaging factor is assumed as 1 and the oscillator level as 0 dBm.
Step 1. Press the
panel once, and the measurement point setup display will apear (Figure 3-4).
Figure 3-4
Measurement point setup display
54
key on the right corner of the front panel to preset the 4287A to its
[Preset]
key on the measurement block (MEASUREMENT) of the front
[Setup View]
Chapter 3