Frequency Reference Adjustment; Aux Input Adjustment; Synthesizer Gain Adjustment - Agilent Technologies E5071C Service Manual

Ena rf network analyzers
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Adjustment
Required Adjustment after Replacing Assembly
Required equipment for frequency reference adjustment
Description
Frequency Counter
Frequency Standard
BNC(m)-BNC(m) Cable, 61 cm

Frequency Reference Adjustment

The purpose of this procedure is to adjust the 10 MHz frequency reference on the source
board.
Required equipment for frequency reference adjustment
Description
Frequency Counter
Frequency Standard
BNC(m)-BNC(m) Cable, 61 cm

AUX Input Adjustment

The purpose of this procedure is to adjust the offset error and proportional error of AUX
Input.
Required equipment for the oven reference adjustment
Description
Function Generator
Multi Meter
BNC(m)-BNC(m) Cable, 61 cm x 4 ea
BNC T Adapter x 2ea
Banana BNC

Synthesizer Gain Adjustment

The purpose of this procedure is to adjust 2Bit ATT of RF Path in Synsthesizer for Source.
Required equipment for the F range gain/phase adjustment
None
172
Recommended Model
Agilent 53181A with Opt. 010 and 124 or Agilent
53131/2A with Opt. 010 and 124
Agilent E5071A
Agilent p/n 8120-1839
Recommended Model
Agilent 53181A with Opt. 010 and 124 or Agilent
53131/2A with Opt. 010 and 124
Agilent 5071A
Agilent p/n 8120-1839
Recommended Model
Agilent 33120A or 33250A
Agilent 34401, 34410, 34411 or 3458A
Agilent p/n 8120-1839
Chapter 3