Performance Tests

2. Source Power Range, Linearity, and Accuracy

2. Source Power Range, Linearity, and Accuracy
Perform this test to verify the power range and linearity at different CW frequencies
throughout the analyzer operating frequency range.
Analyzer warm-up time: 30 minutes
Specifications
Power Range
a
Power Level Accuracy
b
Power Level Linearity
−5 to +15 dBm
+15 to +20 dBm
c
Option 006:
−5 to +13 dBm
+13 to +18 dBm
a. At 25 °C ±5 °C, and at +10 dBm output level.
b. At 25 °C ±5 °C, relative to +10 dBm output level.
c. Only for analyzers with Option 006 (6 GHz frequency
extension).
Required Equipment
Description
Power Meter
Power Sensor
2-way power splitter (2), 50Ω
Attenuator (2), 20 dB
RF cable set, 50Ω type-N
Additional Equipment for Option 006 Analyzers
Power Sensor
2-6
Specification
±1.0 dB
±0.25 dB
±0.5 dB
±0.25 dB
±0.5 dB
HP/Agilent Model Number
436A/437B/438A or E4418B/4419B
8482A
11667A Option 001
8491A Option 020
11851B
8481A