Virtual Bridge Coefficient Adjustment
This item writes the coefficient factors to the E5071C.
Required equipment for the source power adjustment
None
Source Output Power Adjustment
The purpose of this procedure is to adjust the variable attenuator in Level Vernier board,
and the gains among Level Vernier, Distributor and Receiver.
Required equipment for the test port characteristics adjustment
Description
Power Meter
Power Sensor
Receiver IF Range Adjustment
The purpose of this procedure is to adjust IF Range.
Required equipment for the test port characteristics adjustment
Description
Calibration Kit
N(m)-N(m) Cable
Receiver Ports Characteristics Adjustment
The purpose of this procedure is to adjust source match, directivity and tracking.
Required equipment for the test port characteristics adjustment
Description
Calibration Kit
N(m)-N(m) Cable
Receiver Absolute Gain Adjustment
This item writes the coefficient in Absolute measurement.
Required equipment for the test port characteristics adjustment
None
Chapter 3
Required Adjustment after Replacing Assembly
Recommended Model
Agilent E4419A/B
Agilent E9304A Opt H18
Recommended Model
Agilent 85032F
Agilent N6314A (p/n 8120-8862)
Recommended Model
Agilent 85032F
Agilent N6314A (p/n 8120-8862)
Adjustment
173