Virtual Bridge Coefficient Adjustment; Source Output Power Adjustment; Receiver If Range Adjustment; Receiver Ports Characteristics Adjustment - Agilent Technologies E5071C Service Manual

Ena rf network analyzers
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Virtual Bridge Coefficient Adjustment

This item writes the coefficient factors to the E5071C.
Required equipment for the source power adjustment
None

Source Output Power Adjustment

The purpose of this procedure is to adjust the variable attenuator in Level Vernier board,
and the gains among Level Vernier, Distributor and Receiver.
Required equipment for the test port characteristics adjustment
Description
Power Meter
Power Sensor

Receiver IF Range Adjustment

The purpose of this procedure is to adjust IF Range.
Required equipment for the test port characteristics adjustment
Description
Calibration Kit
N(m)-N(m) Cable

Receiver Ports Characteristics Adjustment

The purpose of this procedure is to adjust source match, directivity and tracking.
Required equipment for the test port characteristics adjustment
Description
Calibration Kit
N(m)-N(m) Cable

Receiver Absolute Gain Adjustment

This item writes the coefficient in Absolute measurement.
Required equipment for the test port characteristics adjustment
None
Chapter 3
Required Adjustment after Replacing Assembly
Recommended Model
Agilent E4419A/B
Agilent E9304A Opt H18
Recommended Model
Agilent 85032F
Agilent N6314A (p/n 8120-8862)
Recommended Model
Agilent 85032F
Agilent N6314A (p/n 8120-8862)
Adjustment
173