Making Relative Measurements; Procedure - Agilent Technologies E4418B User Manual

Power meter
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

Power Meter Operation

Making Relative Measurements

Making Relative Measurements
Relative mode allows you to compare any measurement result to a reference value.
The relative reading can be displayed in either dB or % by pressing
selecting either
or
dB
%
multiplier may be shown.
Relative mode is window dependent and when enabled displays "Rel" in the window
to which it's applied.

Procedure

To set a reference value on the currently selected window:
Rel
1.
Press
.
Offset
2.
Press
to use the current reading as the reference value. This allows you
Rel
to compare any measurement result in dB or percentage (%).
3.
Rel Off On
Successive measurements are now displayed relative to the reference value. The
relative mode can be disabled and re-enabled simply by pressing
2-34
. When the measurement result is displayed in % a prefix
is automatically set to "On" when
dBm/W
and
is pressed.
Rel
Rel Off On
Agilent E4418B User's Guide
.