Programming the Analyzer
IEEE Common GPIB Commands
Remote Command
Initial S/W Revision
Recall Instrument State
This command recalls the instrument state from the specified instrument memory register.
• If the state being loaded has a newer firmware revision than the revision of the instrument, no state is
recalled and an error is reported
• If the state being loaded has an equal firmware revision than the revision of the instrument, the state
will be loaded.
• If the state being loaded has an older firmware revision than the revision of the instrument, the
instrument will only load the parts of the state that apply to the older revision.
Remote Command
Example
Notes
Min
Max
Status Bits/OPC dependencies
Initial S/W Revision
Save Instrument State
This command saves the current instrument state and mode to the specified instrument memory register.
Remote Command
Example
Notes
Min
Max
Status Bits/OPC dependencies
Initial S/W Revision
102
*OPT?
Prior to A.02.00
*RCL <register #>
*RCL 7 Recalls the instrument state that is currently stored in register 7.
Registers 0 through 6 are accessible from the front panel in menu keys for
Recall Registers.
0
127
The command is sequential.
Prior to A.02.00
*SAV <register #>
*SAV 9 Saves the instrument state in register 9.
Registers 0 through 6 are accessible from the front panel in menu keys for
Save Registers.
0
127
The command is sequential.
Prior to A.02.00
Chapter 4