Power On; Line Input Receptacle; Fuse; Steps To Turn On The Power - Agilent Technologies 4286A User Manual

Rf lcr meter
Table of Contents

Advertisement

1. Power ON

Verify the connector box and the test head are correctly set up before turning ON the 4286A.
If they have not been set up, see Chapter 2.

Line Input Receptacle

AC Power cable is connected to this receptacle.

Fuse

Use the following fuse:
Agilent Part Number : 2110-0917
(UL/CSA type, Semi Time Lag, 6.3 A 250 V)
If you need this fuse, contact your nearest Agilent Technologies Sales and Service Oce.

Steps to turn on the power

1. Verify the power line setting is correct before you turning ON the meter. If necessary, see
\Changing the Line Voltage Setting" in Appendix B.
2. Connect AC power cable to the line input receptacle.
3. Press the LINE switch.
After the power-on self-test (approximately 10 seconds), the installed options and the
connected test head information are displayed.
Note
Only option numbers 1C2 and 001 are displayed at power-on (if they are
installed). Other installed option numbers are listed on the rear panel.
A 30 minute-warm-up period is required to stabilize the 4286A after it has been turned ON.
This ensures the 4286A its speci ed measurement accuracy.
3-4
Basic Measurement Procedures
Figure 3-3. Line Input Receptacle and Fuse

Advertisement

Table of Contents
loading

Table of Contents