Sampler Magnitude and Phase Correction Constants

(Test 53)

This adjustment procedure corrects the overall flatness of the microwave components that
make up the analyzer receiver and signal separation sections. This is necessary for the
analyzer to meet the published test port flatness specification.
Analyzer warmup time: 30 minutes.
Required Equipment and Tools
Description
Power Meter
Power Sensor
Power Sensor (for Option 006 analyzers)
GPIB Cable
Power splitter
RF cable set (2)
Adapter type-N (m) to type-N (m)
Adapter type-N (f) to type-N (f)
Antistatic Wrist Strap
Antistatic Wrist Strap Cord
Static-control Mat and Ground Wire
Procedure
1. If you just completed
procedure with step 8.
2. Press
Preset
Local
3. Press
SET ADDRESSES
Local
address is 13. Refer to the power meter manual as required to observe or change its
GPIB address.
4. Press
POWER MTR:438A/437
Choose the appropriate model number. (Use the 438A/437 selection if the power meter
is an E4419B or E4418B.)
NOTE
If you are using the 438A or E4419B power meter, connect the 8482A power
sensor to channel A, and the 8481A power sensor to channel B (for adjusting
analyzers with Option 006).
Chapter 3
Sampler Magnitude and Phase Correction Constants (Test 53)
HP/Agilent Part or Model Number
436A/437B/438A or E4418B/4419B
8482A
8481A
10833A
11667A Option 001
11500B
1250-1475
1250-1472
9300-1367
9300-0980
9300-0797
"RF Output Power Correction Constants (Test 47),"
SYSTEM CONTROLLER
ADDRESS: P MTR/GPIB
to toggle between the 438A/437 and 436A power meters.
Adjustments and Correction Constants
.
. The default power meter
continue this
3-19