Avg Mode
Sets the type of termination control used for the averaging function. This determines the
averaging action after the specified number of measurements (see
reached.
Mode
Remote Command
Example
Preset
State Saved
Range
Key Path
Exponential Sets the
Avg Mode
each successive data acquisition is exponentially weighted and combined with the existing
average. This weights new data more than the old data, which facilitates tracking of slow
changing signals.
Example
Key Path
Repeat Sets
Avg Mode
to the repeat function. Repeat averaging shows resulting data after
the selected number of sweeps has been averaged, then clears the data before repeating
the process.
NOTE
With large average counts and/or slow traces, the trace from the last
completed result remains on the display while
58
SA
Front-panel key
SA

[:SENSe]:AVERage:TCONtrol

EXPonential|REPeat|RMAXhold
[:SENSe]:AVERage:TCONtrol?
AVER:TCON REP
Exponential
Saved in instrument state.
Exponential | Repeat | Repeat Max Hold
Meas Setup, Avg Mode
function to exponential. With exponential averaging mode,
AVER:TCON EXP
Meas Setup, Avg Mode, Exponential
Avg Number
) has been
Repeat
averaging is