Power Meter Operation

Editing Frequency Dependent Offset Tables

You can edit frequency dependent offset tables using the "Edit" menu as
shown in
To view the frequency dependent offset tables currently stored in the power
meter, press
displayed as shown in

Figure 2-7 "Edit" Screen

There are ten frequency dependent offset tables named CUSTOM_A through
CUSTOM_J which do not contain any data when the power meter is shipped
from the factory.
You cannot delete any of the 10 existing frequency dependent offset tables or
create any additional tables. However, you can edit the 10 existing tables. Each
frequency dependent offset table can contain a maximum of 80 frequency
points.
To edit a table:
1 Select one using the
60
Figure
2-7.
,
Figure
tables. When the table you want to edit is highlighted, press
"Edit Offset" screen is displayed as shown in
screen you can modify, add, and remove frequencies and offset and also edit
the table name. Use the
table name and its frequencies and offsets.
A frequency in the range of 0.001 MHz to 999.999 GHz can be entered.
,