Editing Frequency Dependent Offset Tables; Figure 2-7 "Edit" Screen - Agilent Technologies E4418B User Manual

Power meter
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

2
Power Meter Operation
If the measurement frequency does not correspond directly to a frequency in the sensor
N O T E
calibration table (if selected) and the frequency dependent offset table being used, the
power meter calculates the calibration factor and offset using linear interpolation.
If you enter a frequency outside the frequency range defined in the sensor calibration table
or the frequency dependent offset table, the power meter uses the highest or lowest
frequency point in the appropriate table to set the calibration factor and offset.
The value of the calibration factor being used by the power meter to make a measurement
is displayed under the

Editing Frequency Dependent Offset Tables

You can edit frequency dependent offset tables using the "Edit" menu as
shown in
To view the frequency dependent offset tables currently stored in the
power meter, press
screen is displayed as shown in

Figure 2-7 "Edit" Screen

56
softkey.
Cal Fac
Figure 2-
7.
,
,
Tables Freq Dep Offset
Figure 2-
6.
E4418B Power Meter User's Guide
. The "Offset Tbls"