Editing Frequency Dependent Offset Tables - Agilent Technologies N1913a User Manual

Epm series power meters
Hide thumbs Also See for N1913a:
Table of Contents

Advertisement

2
General Power Meter Functions

Editing Frequency Dependent Offset Tables

There are ten frequency dependent offset tables named CUSTOM_A through
CUSTOM_J. They are empty of any data when the power meter is shipped
from the factory.
You cannot delete any of the 10 existing frequency dependent offset tables
or create any additional tables. However, you can enter values into the 10
existing tables. Each frequency dependent offset table can contain a
maximum of 80 frequency points.
To view the frequency dependent offset tables currently stored in the
power meter, press
Dependent Offset Tables screen is displayed as shown in
Editing frequency dependent offset tables requires the following steps:
1 Identify and select the table you want to edit
2 Rename the table
3 Enter the frequency and offset pairs
4 Save the table
Procedure
First, select the table you want to edit as follows:
A frequency in the range of 0.001 MHz to 1000.0 GHz can be entered. A calibration factor in
N O T E
the range of 0.0000000001 to 10000000000 can be entered.
1 Press
2 Choose the table you want to edit using the
44
Tables
,
Edit Table
to display the Edit Offset screen as shown in
,
,
Tables
Freq. Dep. Offset
Freq. Dep. Offset
to display the Offset Tbls screen.
,
N1913/1914A EPM Series Power Meters User's Guide
. The Frequency
Figure 2-
8.
and
keys. Press
Figure 2-
10.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for: