Function Specific Troubleshooting - Agilent Technologies ENA Series Service Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Troubleshooting

Function Specific Troubleshooting

Function Specific Troubleshooting
If the E5061A/E5062A exhibits a failure symptom that is related to a specific function or
control such as a front panel key control, display, data storage, remote control interface,
printer interface, external trigger, external keyboard or mouse, isolate the trouble using the
Function Specific Troubleshooting procedures described below. The major functions of the
E5061A/E5062A and the troubleshooting procedure for each function are shown in Table
4-4.
Table 4-4
Major functions and troubleshooting procedures
Function
Front panel keys
Touch panel
LCD display
External keyboard
Mouse
Floppy disk drive
Video output
External trigger input
106
Description
All the E5061A/E5062A functions except for
VBA and service functions can be set and
controlled via the front panel keys.
The E5061A/E5062A equipped with option 016
has a touch screen display that allows all the
functions in the menu bars, setup windows and
dialog boxes to be set by a touch to the screen
panel.
Almost all the information including the
measurement value, setup state, result data
processing, menu bar, softkey label and others are
indicated on the 10.4-inch color LCD display.
The external keyboard can be used for the entry of
numerical and character data when it is connected
to the keyboard interface connector (PS-2) on the
rear panel.
The mouse can be used to move the pointer on the
LCD display, select a function and change a
setting, when it is connected to the mouse interface
connector (PS-2) on the rear panel.
The 3.5-inch floppy disk drive on the front panel is
used to save the E5061A/E5062A setup state and
measurement data in a 3.5-inch, 1.4 MB floppy
disk in MS-DOS compatible format.
An external color monitor can be used to display
the same information as the E5061A/E5062A
LCD display, when it is connected to the Video
output connector (24-pin D-Sub) on the rear panel.
The external trigger input terminal (BNC) on the
rear panel allows an external trigger source to be
used for measurement trigger.
Troubleshooting
Refer to "To Check the Front
Panel" on page 109.
Refer to "To Check the
Touch Panel (Option 016
only)" on page 110.
Refer to "To Check the
LCD" on page 110.
Refer to "To Check the
External Keyboard" on
page 111.
Refer to "To Check the
Mouse" on page 111.
Refer to "To Check the
FDD" on page 111.
Refer to "To Check the
Video output" on page 112.
Refer to "To Check the
External Trigger Input" on
page 112.
Chapter 4

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5061aE5062a

Table of Contents