4
Calibrating

Calibration Strategy

62
The 16962A logic analyzer does not require any periodic
adjustments or calibration by the user to ensure operational
accuracy.
However, Agilent recommends that performance of the
16962A logic analyzer be tested against its specifications at
two- year intervals (see
Performance"
on page 21). This testing is required in order
to obtain calibration certification.
"Testing Logic Analyzer
16962A Logic Analyzer Service Guide