Troubleshooting the Logic Analyzer

Error Messages

Error Messages
This section lists some of the messages that the analyzer displays when
it encounters a problem.

". . . Inverse Assembler Not Found"

This error occurs if you rename or delete the inverse assembler file
that is attached to the configuration file. Ensure that the inverse
assembler file is not renamed or deleted, and that it is on the same
flexible disk or in the same directory as the configuration file.

"No Configuration File Loaded"

This is usually caused by trying to load a configuration file for one type
of module or the system into a different type of module.
Verify that the appropriate module has been selected as the target of
the Load operation. Selecting Load All will cause incorrect operation
when loading most analysis probe configuration files.
See Also
"To Load a Configuration" on page 242.