Recall Now
Initiates the recall of a saved state or trace data file depending on the file type currently
selected. When pressed a screen is displayed that shows a list of saved files. Using the up
and down arrow navigation keys, you must highlight one of the file names and press either
Select
the
navigation key or the
recalled into the analyzer. When traces are recalled, they always recall in View mode.
Traces saved in TRC format can be recalled individually. When a trace is recalled, the state
that existed when that trace was saved is recalled along with the trace. In addition, the
sweep is set to single.
NOTE
To compare two saved traces from different saves, place traces in view
mode before saving them. This prevents the trace from being rewritten
based on a state change from subsequent recalls.
Mode
Remote Command
Example
Chapter 4
Select
menu key. The selected file data will then be
SA, Stimulus/Response
:MMEMory:LOAD:STATe 1,<filename>
:MMEMory:LOAD:TRACe <trace>,<file_name>
*RCL <register>
MMEM:LOAD:STAT 1,'MyState.sta' recalls the saved
instrument state file name 'MyState.sta' into the analyzer. The
.sta extension is required.
*RCL 12 recalls the instrument state from the instrument
memory register REG12. The available register numbers are 0 to
127.
MMEM:LOAD:TRAC TRACE3,'MyTrace.trc' recalls the trace +
state file MyTrace.trc into the analyzer as trace 3.