Opt; Rcl; Rst - Agilent Technologies 4288A Programming Manual

1khz/1mhz capacitance meter
Table of Contents

Advertisement

Command Reference

*OPT?

*OPT?
Syntax
*OPT?
Description
Reads out the identification number of an option installed in the 4288A. (Query only)
Query response
{numeric}<newline><^END>
If there is no installed option, 0 is read out.
Equivalent key
[Config]
sequence

*RCL

Syntax
*RCL <numeric>
Description
Recalls the instrument setups saved in the register of the specified number on EEPROM.
For details of recalled instrument setups, refer to Appendix D, "Initial Settings." (No
query)
Parameters
If the specified parameter is out of the allowable setup range, an error occurs.
Related commands *SAV on page 127
Equivalent key
[Rcl]
sequence

*RST

Syntax
*RST
Description
Resets the instrument to the preset state.
The preset state is different from that when resetting is done using the :SYST:PRES
command. For details, refer to Appendix D, "Initial Settings." (No query)
Related commands :SYST:PRES on page 193
:INIT:CONT on page 183
Equivalent key
No equivalent key is available on the front panel.
sequence
126
(
-
) -
[Blue]
[-]
Rev
Description
Specified number
Range
0 to 9
Resolution
1
<numeric>
Chapter 10