Making Measurements
AM Depth Measurement
Default is
.
Free Run
is selected, the default trigger level is − 6 dBm. To
When trigger source
Video
adjust, press the
hardkey, then
.
Trig
Level
Capture Time
Requires PSA firmware revision A.11.08 or above, or Option 23A
Sets the data capture time. Manual and automatic modes are available.
Default is 250 ms.
Fast Mode
When fast mode is on, PSA samples fewer points so the capture time and
processing time is shorter. This results in faster measurements. The speed
improvement is dependent on signal rate and deviation. Accuracy may be affected
and the instrument specifications are given with Fast mode Off.
Default is OFF
AM Depth Only
Allows you to select only the AM Depth results. The results for Modulation Rate,
Distortion and SINAD are displayed as "---". Measurement speed increases when
only AM Depth results are returned. When No is selected, measurement results for
AM Depth, Modulation Rate, Modulation Distortion and Modulation SINAD are
provided.

Adjusting the Detector, and High and Low Pass Filters

Press
hard key.
Det/Demod
Detector
Peak −
Selecting the
or
allows you to check modulation symmetry. For
Peak +
Peak −
asymmetrical baseband waveforms, selecting
or
will result in
Peak +
different measurement results.
Peak+-/2
Displays the average value of the positive and negative peak readings.
RMS
Selects the average detector.
Peak Hold
Functional only when the detector is currently set to Peak + or Peak −. Holds the
maximum peak deviation peak positive or peak negative value from each time
measurement. Default is Off.
Chapter 3
84