Executing Low-Loss Capacitor Calibration - Agilent Technologies Twelfth 4287A RF LCR Operation Manual

Table of Contents

Advertisement

Learning Basic Operations
Calibration and Compensation
Step 2. While pressing the open standard (0S) from above, turn it clockwise and securely connect
it to the 7-mm terminal (Figure 3-16).
Figure 3-16
How to connect open standard
Step 3. Press the
cursor to the
Step 4. Press the
in the Calibration progress indicator in the calibration window indicates the progress of the
open calibration data measurement. Upon completion of open calibration data
measurement, the screen changes to the next window (Figure 3-17).

Executing low-loss capacitor calibration

Figure 3-17
Low-loss capacitor calibration window
Step 1. Turn the outside of the open standard counterclockwise and remove it from the 7-mm
terminal.
Step 2. While pressing the low-loss capacitor from above, turn it clockwise and securely connect it
to the 7-mm terminal (Figure 3-18).
62
key or
key on the front panel the required number of times to move the
Next>
button.
key on the front panel to start measurement of open calibration data. The bar
Chapter 3