Performance Test

5. TRACE NOISE TEST

5. TRACE NOISE TEST

Description

This test checks the trace noise level for each test port of E5071C. The trace noise level is
quantified by performing a "through" measurement 32 times from 9 kHz to 8.5 GHz range,
with a cable connected between two ports. The trace noise level for S11, S22, S33 and S44
are quantified by performing a "short" measurement 32 times from 9 kHz to 8.5 GHz
range, with a cable (terminated with "Short" device) connected to each ports(1,2,3, and 4).
Standard deviation of the measured values at each frequency is calculated and, then
translated into a noise level expressed in dB rms / deg rms.

Test equipment

20 inch 50 ohm cable
N-SMA Adapter
Type N, 50 ohm Short termination (1ea)
34
Agilent p/n 5062-6691
Agilent p/n 1250-2879
Agilent p/n 85032-60009